What is analyser interfacing in a lab?
Every modern analyser produces results electronically. Without interfacing, a technician reads them from the screen or printout and types them into the software – slow, and a frequent source of errors such as swapped digits or values entered for the wrong patient. Interfacing lets the analyser send results directly to the LIS, matched by sample barcode or ID.
What are ASTM and HL7?
| Protocol | Typical use | Notes |
|---|---|---|
| ASTM (E1381/E1394, now CLSI LIS1/LIS2) | Most haematology, biochemistry, immunoassay and electrolyte analysers | Record-based messages (header, patient, order, result); works over serial RS-232 or TCP/IP |
| HL7 v2 | Newer analysers, middleware and hospital systems | Segment-based messages (MSH, PID, OBR, OBX); usually TCP/IP |
| Vendor formats | Some small or older machines | Custom text output needing a specific driver |
QLab interfaces analysers over ASTM; see machine interfacing.
What is the difference between unidirectional and bidirectional interfacing?
- Unidirectional – the analyser sends results to the LIS. The technician still selects tests on the machine.
- Bidirectional – the LIS also sends the test orders to the analyser (often when it scans the barcode), so the machine knows what to run. Saves more time but needs analyser support and more setup.
How is an analyser connected?
- Check the analyser's LIS settings and host interface manual (protocol, baud rate, port or IP).
- Connect by serial cable (often with a USB-serial adapter) or by network to the interfacing computer.
- Map each analyser parameter code to the matching test parameter in the LIS.
- Run test samples and compare values on both screens.
- Go live, keeping manual entry available as a fallback.
What should you ask your LIS vendor about interfacing?
- Is my exact analyser model already supported?
- Is interfacing included or charged per machine?
- Do results still require technician review and pathologist approval before release? (They should.)
- What happens to results if the interfacing PC is off?
- Does it work with the offline version as well as the cloud version?
Ask these during a QLab demo.
Is interfacing worth it for a small lab?
If you run a 3-part or 5-part cell counter, interfacing is usually the first to pay off: a CBC has 15–25 parameters, so typing each one is slow and error-prone. Biochemistry analysers with many tests per sample follow. Labs doing mostly manual tests (serology cards, urine microscopy) gain less and can keep using fast manual result entry.
Which analysers are commonly interfaced in Indian labs?
| Analyser type | Why interface it |
|---|---|
| 3-part / 5-part haematology counter | 15–25 parameters per CBC; the highest typing load |
| Fully automatic biochemistry analyser | Many tests per sample, high daily volume |
| Electrolyte analyser | Several values per run, frequent repeats |
| Immunoassay / hormone analyser | Precise numeric values where typos matter |
| HbA1c analyser | Steady diabetes-driven volume |
| Urine analyser | Strip parameters captured by machine |
What does an ASTM message contain?
An ASTM transmission is a sequence of records, each starting with a letter:
- H – header: sender, analyser model, date and time.
- P – patient record (often minimal; the sample ID is what matters).
- O – order: sample ID or barcode and requested tests.
- R – result: test code, value, unit, flags.
- L – terminator: end of message.
The LIS reads the sample ID from the O record, finds the matching bill and places each R value in the mapped test parameter. Low-level framing (checksums and ENQ/ACK handshakes) makes sure nothing is lost on the cable.
What goes wrong during interfacing, and how is it fixed?
- No data received – wrong baud rate, cable type (straight vs null-modem) or port; check settings on both sides.
- Results not matching patients – the analyser uses its own sequence number instead of the barcode; enable barcode reading.
- Some parameters missing – incomplete code mapping; map every parameter the analyser sends.
- Units differ – align units in the test master with the analyser output.
Good vendors handle these during setup; ask about your models in a demo.
Frequently asked questions
Which protocol do most lab analysers use?
Most clinical analysers support ASTM over serial or TCP/IP; some newer ones support HL7.
Does QLab support machine interfacing?
Yes. QLab connects analysers over ASTM so results flow in without typing, then go through review and approval.
Do interfaced results need approval?
Yes. Interfaced values should still be reviewed and approved by an authorised person before the report is released.
Can a semi-auto analyser be interfaced?
Only if it has a data output port and documented host protocol. Many semi-auto models do not, so results are entered manually.